CVE-2019-25724

6.5 MEDIUM
Published: June 02, 2026 Modified: June 04, 2026
View on NVD

Description

DrΓ€ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

AI Explanation

Get an AI-powered plain-language explanation of this vulnerability and remediation steps.

Login to generate AI explanation

CVSS v3.x Details

0.0 Low Medium High Critical 10.0
Vector String
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

References to Advisories, Solutions, and Tools

Patch Vendor Advisory Exploit Third Party Advisory

2 reference(s) from NVD

Quick Stats

CVSS v3 Score
6.5 / 10.0
EPSS (Exploit Probability)
0.2%
7th percentile
Exploitation Status
Not in CISA KEV

Weaknesses (CWE)