CVE-2020-16220

4.3 MEDIUM
Published: September 11, 2020 Modified: February 23, 2026
View on NVD

Description

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.

AI Explanation

Get an AI-powered plain-language explanation of this vulnerability and remediation steps.

Login to generate AI explanation

CVSS v3.x Details

0.0 Low Medium High Critical 10.0
Vector String
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L

References to Advisories, Solutions, and Tools

Patch Vendor Advisory Exploit Third Party Advisory
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01
Source: ics-cert@hq.dhs.gov
Third Party Advisory US Government Resource
https://www.philips.com/productsecurity
Source: ics-cert@hq.dhs.gov
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party Advisory US Government Resource
https://www.philips.com/productsecurity
Source: af854a3a-2127-422b-91ae-364da2661108

4 reference(s) from NVD

Quick Stats

CVSS v3 Score
4.3 / 10.0
EPSS (Exploit Probability)
0.0%
10th percentile
Exploitation Status
Not in CISA KEV

Weaknesses (CWE)

Affected Vendors

philips