CVE-2025-54304

9.8 CRITICAL
Published: December 04, 2025 Modified: December 16, 2025

Description

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

AI Explanation

Get an AI-powered plain-language explanation of this vulnerability and remediation steps.

Login to generate AI explanation

CVSS v3.x Details

0.0 Low Medium High Critical 10.0
Vector String
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

References to Advisories, Solutions, and Tools

Patch Vendor Advisory Exploit Third Party Advisory

3 reference(s) from NVD

Quick Stats

CVSS v3 Score
9.8 / 10.0
EPSS (Exploit Probability)
0.1%
26th percentile
Exploitation Status
Not in CISA KEV

Weaknesses (CWE)

Affected Vendors

thermofisher